Physio-control LIFEPAK 1000 Manuel d'utilisateur Page 51

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 86
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 50
Specifications
LIFEPAK 1000 Defibrillator Operating Instructions A-1
©2006-2012 Physio-Control, Inc.
SPECIFICATIONS
All specifications are at 20°C (68°F) unless otherwise stated.
Defibrillator
Waveform Biphasic Truncated Exponential with voltage and duration
compensation for patient impedance.
With Adult Pads:
Patient Impedance Range: 10 – 300 ohms
The following specifications apply from 25 to 175 ohms.
Energy Accuracy:
10% of the energy setting into 50 ohms
15% of the rated energy output into 25 – 175 ohms
Rated energy output is the nominal delivered energy based on the
energy setting and patient impedance, as defined in the following chart.
* Energy setting selected
Rate d Energy O utput
125
150
175
200
225
250
275
300
325
350
375
25 50 75 100 125 150 175
Patient Im pedance (ohm s)
Energy (
J
325 J *
300 J *
275 J *
250 J *
225 J *
200 J *
175 J *
150 J *
360 J *
Rated Energy Output
Energy (J)
Patient Impedance ()
Vue de la page 50