3-6 LIFEPAK 1000 Defibrillator Operating Instructions
How to Use the LIFEPAK 1000 Defibrillator
DEFIBRILLATION IN MANUAL MODE
The LIFEPAK 1000 defibrillator provides a Manual mode to override the AED features of the
defibrillator. Manual mode provides operator-initiated analysis, charge, shock, and disarm
functions. This mode is useful in a tiered response system when a provider trained in manual
defibrillation and authorized to place the defibrillator in Manual mode takes over the scene from
a BLS-AED trained provider.
To use Manual mode:
1 Press the Menu button.
2Select
YES to enter Manual mode. The ECG trace and Heart Rate Indicator appear on
the screen.
3 If the displayed ECG rhythm appears shockable, press
CHARGE to initiate charging of
the defibrillator. The screen will indicate that the defibrillator is charging and a charge tone
will sound.
4 Clear everyone away from the patient, bed, or any equipment connected to the patient.
5 When the charge is complete, press the flashing (shock) button to deliver energy to
the patient.
6 After delivering a shock, the energy for each subsequent shock is automatically selected
based on the energy level configured in Setup.
Note: To remove an unwanted charge at any time, press
DISARM.
Analysis
The LIFEPAK 1000 defibrillator can be set up to display an ANALYZE softkey when in
Manual mode.
To initiate an analysis:
1 Confirm that the patient is unresponsive, not breathing, and without a pulse.
2 Press
ANALYZE.
3 If the rhythm analysis results in a No Shock Advised decision, the defibrillator remains in
Manual mode without further prompts.
4 If the rhythm analysis results in a Shock Advised decision, the defibrillator automatically
begins charging accompanied by a charge tone. If you determine that a shock is not
warranted, press
DISARM.
5 When the charge is complete, clear everyone away from the patient, bed, or any equipment
connected to the patient.
6 Press the flashing (shock) button to deliver energy to the patient.
7 After delivering a shock, the defibrillator remains in Manual mode.